Preliminary test results of LAr prototype detector | Pei-Xian Li
; Meng-Yun Guan
; Chang-Gen Yang
; Peng Zhang
; Jin-Chang Liu
; Yong-Pen Zhang
; Cong Guo
; Yi Wang
; | Date: |
Wed, 6 Jan 2016 04:34:18 GMT (1853kb) | Abstract: | WIMPs are a well-motivated galactic dark matter candidate. Liquid argon (LAr)
is an attractive target for the direct detection of WIMPs. The LAr prototype
detector is designed to study the technology and property of LAr detector. The
prototype detector have an active volume containing 0.65 kg of liquid argon.
The liquid nitrogen(LN) cooling system allows the temperature of liquid argon
to be maintained at the boiling point (87.8 K) with fluctuations less than 0.1
K. The prototype was calibrated with a Na$^{22}$ source, with the light yield
1.591$pm$0.019 p.e./keV for the 511 keV gamma rays using the domestic-made
argon purification system. | Source: | arXiv, 1601.1075 | Services: | Forum | Review | PDF | Favorites |
|
|