Characterising a Si(Li) detector element for the SIXA X-ray spectrometer | T. Tikkanen
; S. Kraft
; F. Scholze
; R. Thornagel
; G. Ulm
; | Date: |
4 Mar 1997 | Journal: | Nucl. Instrum. Meth. A390 (1997) 329 | Subject: | Instrumentation and Detectors | physics.ins-det | Affiliation: | Observatory, University of Helsinki), S. Kraft, F. Scholze, R. Thornagel, G. Ulm (Physikalisch-Technische Bundesanstalt, Berlin | Abstract: | The detection efficiency and response function of a Si(Li) detector element for the SIXA spectrometer have been determined in the 500 eV to 5 keV energy range using synchrotron radiation emitted at a bending magnet of the electron storage ring BESSY, which is a primary radiation standard. The agreement between the measured spectrum and the model calculation is better than 2%. PACS: 95.55.Ka; 07.85.Nc; 29.40.Wk; 85.30.De Keywords: Si(Li) detectors, X-ray spectrometers, detector calibration, X-ray response, spectral lineshape | Source: | arXiv, physics/9703004 | Services: | Forum | Review | PDF | Favorites |
|
|